SPINAL DURAL ARTERIOVENOUS FÍSTULA: INDOCYANINE GREEN VIDEOANGIOGRAPHY IN NEGATIVE

. 2015 Aug; 5(4): e5–e6

Simal Julián et al.

2013 May;22 Suppl 3:S471-7.

Simal Julian et al

www.neurocirugiasimal.com

Tabla de contenidos

PURPOSE

Indocyanine green videoangiography (IGV) has proven its effectiveness in the field of exovascular neurosurgery, both in the intracranial and spinal compartment, but is necessary to define a systematic process for the performance of the IGV to facilitate its interpretation during the procedure. We have defined and applied the concept of videoangiography «in negative» (INIGV) to spinal dural arteriovenous fistulae (dAVF) for the detection and treatment of arteriovenous shunts, so called because the first phase is performed with the vessel suggestive of being pathological occluded.

METHODS

A Pentero-operating microscope with near-infrared IGV-integrated system (Carl Zeiss Co., Germany) was used. At our institution, 24 patients were treated for a spinal dAVF between 1995 and 2011, only in the last 4 cases, INIGV was performed.

RESULTS

We describe the IGV in negative procedure and show the most illustrative cases. In all cases, the fistula occlusion was confirmed by postoperative selective digital subtraction angiography (DSA). INIGV demonstrate its capacity in detecting vessels not actually arterialized that should be respected and avoid some of the main limitations of the conventional IGV. This is a technical description about an Indocyanine green (ICG) videoangiographic procedure modification that is superior to merely performing ICG before and after clipping of a dAVF.

CONCLUSION

The INIGV results are rapid and easy to interpret procedure and provide great advantages to the dAVF treatment. Nevertheless, further studies are needed with a larger sample size to determine if INIGV may reduce the need to perform immediate postoperative DSA.